On 5 November 2026, the National Medical Products Administration released Notice No. 38 of 2026 on unannounced inspections of five medical device manufacturers. The notice states that serious deficiencies were found in the quality management systems of the five companies.
Hunan Aipute Medical Devices Co., Ltd. added an automated production line for its hydrophilic coated guidewire (registration No. Hunan Medical Device Registration Approval 20252030122), but did not fully conduct risk assessment for changes in key raw materials and production processes. The new automated line moved the key “hydrophilic coating” process from a 10,000-class clean area to a non-clean area without fully assessing the risks introduced by the change. Some key processes lacked validation records, some work instructions for key processes could not guide actual production, and the process inspection procedures were inconsistent with the production flow chart. These circumstances did not comply with the relevant requirements of the Medical Device Production Quality Management Specification and the Appendix for Sterile Medical Devices concerning premises and facilities, design and development, production management, and quality control.
Shenzhen Bochuang Dental Technology Co., Ltd. did not clarify the control methods for some performance indicators in the product technical requirements, did not carry out market release in accordance with the contract manufacturing quality agreement, and could not provide inspection reports or certificates of conformity for some raw materials. These circumstances did not comply with the relevant requirements of the Specification and the Appendix for Customized Dentures concerning procurement and quality control.
Guizhou Qingchunquan Cell Biotechnology Co., Ltd., as a medical device registration holder, currently has no actual office premises and has not equipped sufficient full-time quality management personnel with adequate competence, or technical personnel familiar with the product and with corresponding professional knowledge, and is unable to effectively monitor and control contract manufacturing activities. Some key processes were not validated, and the suitability validation for radiation sterilization was insufficient. These circumstances did not comply with the relevant provisions of the NMPA Announcement on Further Strengthening the Supervision and Administration of Contract Manufacturing by Medical Device Registration Holders (No. 38 of 2024) and the relevant requirements of the Specification concerning production management.
Guizhou Chuanding Biotechnology Co., Ltd. could not provide validation reports for some key processes, and the suitability validation for radiation sterilization was insufficient. The quantity released for production exceeded the actual production quantity; the radiation sterilization certificate did not reflect the product quantity and production batch number; and production records did not meet traceability requirements. The company could not provide records for the preparation of culture media for sterility testing or sterility test records, and inspection records did not meet traceability requirements. These circumstances did not comply with the relevant requirements of the Specification concerning production management and quality control.
Weihai Kaiying Medical Co., Ltd., as a medical device registration holder, has closed and ceased use of its domicile, and relevant staff have been dismissed. It is unable to perform its main responsibility for medical device quality and safety. These circumstances did not comply with the relevant provisions of the NMPA Announcement on Further Strengthening the Supervision and Administration of Contract Manufacturing by Medical Device Registration Holders (No. 38 of 2024).
The five companies have confirmed the identified problems. Provincial drug regulatory authorities shall, in accordance with Article 72 of the Regulations on the Supervision and Administration of Medical Devices, take control measures to order suspension of production against the automated production line for hydrophilic coated guidewires (registration No. Hunan Medical Device Registration Approval 20252030122) of Hunan Aipute Medical Devices Co., Ltd., and against Shenzhen Bochuang Dental Technology Co., Ltd., Guizhou Qingchunquan Cell Biotechnology Co., Ltd., Guizhou Chuanding Biotechnology Co., Ltd., and Weihai Kaiying Medical Co., Ltd., and conduct targeted supervision sampling where necessary. Where suspected violations of the Regulations on the Supervision and Administration of Medical Devices and related provisions are found, they shall be handled strictly in accordance with the law. The companies shall be ordered to assess product safety risks and, where potential safety hazards may exist, recall relevant products in accordance with the Measures for the Administration of Medical Device Recalls. Production may resume only after the companies complete rectification of all deficiency items and pass re-inspection by the provincial drug regulatory authority.
The regulatory signal is clear: registration holder responsibilities, contract manufacturing management, key process validation, clean area change control, release, and traceability records are all priorities in unannounced inspections. For companies, a quality system cannot remain at the document level, and change management cannot be handled as a “do first, document later” exercise.