An NMPA flight inspection notice has once again brought quality management system deficiencies in medical device manufacturing into focus. On 19 November 2026, the National Medical Products Administration issued Notice No. 41 of 2026, disclosing the results of unannounced inspections at Tianjin Zhixin Hongda Medical Device Development Co., Ltd. and Hunan Kangcheng Medical Technology Co., Ltd. Both companies were found to have serious deficiencies in their quality management systems.
At Tianjin Zhixin Hongda Medical Device Development Co., Ltd., the problem centered on a substantive failure in testing. Inspectors found that testing personnel had not actually performed the citrate ion concentration test, and that testing records had been modified, deleted, copied, and fabricated. This does not meet the quality control requirements of the Medical Device Production Quality Management Practices. The company’s registered products involve hemodialysis dry powder and hemodialysis concentrate, which are used directly in dialysis treatment for patients with renal failure. The authenticity of testing records and the actual execution of testing activities are key safeguards before product release. When testing data can be fabricated, the “compliance” on which patients rely loses its factual basis.
At Hunan Kangcheng Medical Technology Co., Ltd., the deficiencies exposed simultaneous gaps at both the documentary and operational levels of sterility assurance. The company’s release testing procedures did not specify media suitability checks, sterility tests, sensitivity checks, or method suitability tests for sterility testing items, and some testing items were not carried out in accordance with the release testing procedures. In addition, expired raw materials stored in the warehouse were not controlled as nonconforming products, failing to meet the quality control and nonconforming product control requirements of the Practices. The company’s main product is a disposable trocar, used in laparoscopic surgery to establish a channel for endoscopes and instruments. Sterility is a precondition for the safe use of such products. The combination of procedural gaps and operational deviations left sterility release without a verifiable technical basis.
The notice confirms that both companies have acknowledged the identified problems. Provincial-level drug regulatory authorities will, in accordance with Article 72 of the Regulations on the Supervision and Administration of Medical Devices, take control measures to order suspension of production, conduct supervisory sampling where necessary, and seriously handle suspected violations of the Regulations and relevant provisions. They will also require the companies to assess product safety risks and, where potential hazards exist, recall relevant products in accordance with the Measures for the Administration of Medical Device Recalls. Production may resume only after all deficiency items have been rectified and re-inspection by the provincial regulatory authority has confirmed compliance.
The deterrent force of flight inspections lies not in the inspection format itself, but in the structural risk it repeatedly reveals. When testing is treated as optional rather than mandatory, and when records are treated as editable documents rather than mirrors of fact, the quality management system degrades from a safety barrier into a shell of paper compliance. For medical device manufacturers, the real starting point of remediation may not be filling in a missing testing procedure or a missing record, but reaffirming a premise: a production license carries the trust patients place in their own bodies.