On November 25, 2026, the Center for Medical Device Evaluation of NMPA released a draft guidance on application requirements for single-arm performance criterion trials, with comments due by December 18, 2026. The document refines requirements under the existing medical device clinical trial design framework, does not create administrative approval requirements, is not enforceable as regulation, and does not apply to in vitro diagnostics managed as medical devices.
A single-arm performance criterion trial assigns all subjects to one intervention, omits a concurrent parallel control, and compares the primary endpoint result with a pre-specified clinically meaningful performance goal. The draft states clearly that evidence from this design ranks lower than a randomized controlled trial and cannot replace an RCT. It is a constrained option, not a default pathway.
Applicability is assessed on three fronts, including whether the device technology is mature and the disease well understood, or whether a control is objectively infeasible; whether the population and endpoints can be sufficiently and objectively defined and remain relatively stable; and whether a recognized performance goal can be obtained or scientifically constructed. The draft draws hard lines through negative examples. A novel mechanism of action, a new indication, or the absence of a comparable predicate does not justify single-arm design on grounds of infeasible control. Nor do a high predicate price or a site’s lack of procurement qualify as objective infeasibility. An interventional left ventricular assist device for high-risk PCI support is cited as a case in which current evidence does not support single-arm design and a randomized comparison is required.
Performance goal construction is the technical centerpiece. The draft distinguishes Objective Performance Criteria (OPC) issued by authoritative bodies from Performance Goals (PG) constructed from evidence. OPC carries greater acceptance and evidentiary weight. Where OPC is unavailable and a PG is used, the overall evidence level of the single-arm trial drops further. PG construction should use methods such as meta-analysis, draw on data reflecting current clinical practice, cover the proposed indication, and demonstrate adequate precision. If the 95% confidence interval is too wide, a trial result that technically meets the goal still cannot establish that the device reaches the accepted minimum standard in the field.
Bias control spans design, conduct, and analysis. The draft catalogs selection bias, confounding, measurement bias, missing-data bias, performance bias, reporting bias, and others. It requires pre-specified endpoint adjudication procedures and a statistical analysis plan, an independent blinded endpoint committee that does not know treatment assignment, and complete documentation of screening and loss to follow-up. External-control adjustment using data from the performance goal source is permitted only as a sensitivity analysis and cannot address unmeasured confounding.
Comments can be sent to tengyy@cmde.org.cn by December 18, 2026. For sponsors considering a single-arm performance criterion path, the signal is clear. The applicability bar is rising, the evidentiary burden for the performance goal is increasing, and bias control cannot be satisfied by a line in the protocol stating that blinding will be attempted.